20
Installation Note E8364-90019
Step 8. Reinstall the Front Panel Assembly and Install Front Panel Jumpers
Refer to
for this procedure.
1. Lift and support the front of the analyzer chassis..
2. Tighten all 12 of the front-panel feed-through connectors using a 5/16-inch torque wrench set to 21-in lbs.
3. Reconnect the ribbon cable (item
{
) to the A3 front panel interface board.
4. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
cable (item
{
) is located below the fan to prevent it from being damaged by the fan blades.
5. With a T-10 TORX driver, install the eight screws (item
|
) in the sides of the frame.
6. Install the six semirigid jumpers (item
}
) on the front panel, and tighten to 10-in lbs.
Figure 13 Front Panel Assembly Reinstallation
NOTE
The previous figure shows the E8363/4B front panel and floppy disk drive.
The E8363/4C front panel has a slightly different appearance and does not
include a floppy disk drive.