Service Guide E8364-90001
7- 65
PNA Series Microwave Network Analyzers
Repair and Replacement Procedures
E8362A, E8363A, E8364A
Removing and Replacing the A41 Hard Disk Drive
c. Adjust the horizontal width of the
Windows
Explorer
window to approximately
one-half the width of the display and position it to either the left or right side of the
display.
d. Repeat this procedure to create another copy of the
Windows
Explorer
window and
position it to the opposite side of the display so that the two copies are displayed
side-by-side.
3. Set one copy of
Windows
Explorer
to the
A:
drive and the other to the
C
: drive:
a. In one copy of
Windows
Explorer
, double-click on
My
Computer
, and then
double-click on
3 1/2 Floppy (A:)
. A listing of the files on the floppy disk will appear
in the right-hand side of the window.
b. In the other copy of
Windows
Explorer
, double-click on
My
Computer
, and then
double-click on
Hard Disk (C:)
.
c. Navigate to
C:\Program Files\Agilent\Network Analyzer
by double-clicking on
each of the names in this path name, in turn. A listing of the files in this directory on
the hard disk will appear in the right-hand side of the window.
4. Copy each of the backup files from the floppy disk to the hard disk drive:
a. Click and drag each of the files from the
A:
drive to the
C:\Program
Files\Agilent \Network Analyzer
directory by clicking and holding on the file
name in the
A:
directory and, while holding the mouse button down, drag the file
name to the
Network Analyzer
directory on the
C:
drive. If you need help with this
“dragging” process, refer to Windows Help. (Click
START
,
HELP
. Click on the
Index
tab and type
dragging
in the dialog box.)
5. After all files have been copied, remove the floppy disk from the analyzer’s disk drive
and exit both copies of
Windows Explorer
.
Set Up the Agilent Administrator Password
If the hard disk has been replaced, it is necessary to set up the Agilent administrator
password for the analyzer. This requires obtaining a password from Agilent:
Obtaining a Password
• If this procedure is being performed by Agilent personnel, go to the following Internet
site and follow the instructions under “Agilent Administrator Password.”
http://mktwww.soco.agilent.com/field/service/network/pna/support.htm
• If this procedure is being performed by non-Agilent personnel, contact Agilent for a
password before beginning this procedure. Refer to
“Contacting Agilent” on page 2-9
Entering Your Agilent Administrator Password
The network analyzer must be powered up and operating for this procedure.
1. Exit the Network Analyzer program, if it is running.
2. On the
Windows Desktop
, right-click on
My Computer
, and then click
Manage
.
3. In the
Computer Management
window, navigate to
System Tools
,
Local Users and
Groups
,
Users
.
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Страница 302: ...A 16 Service Guide E8364 90001 Error Terms PNA Series Microwave Network Analyzers Error Term Data E8362A E8363A E8364A ...
Страница 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Страница 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Страница 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...