Service Guide E8364-90001
4- 51
PNA Series Microwave Network Analyzers
Troubleshooting
E8362A, E8363A, E8364A
Measurement System Troubleshooting
3. Connect the spectrum analyzer to the suspect receiver connector.
4. The measured signal on the spectrum analyzer should be at 41.667 kHz.
• If the measured signal is present, replace the A6 SPAM board. Refer to
and Replacing the A6, A8, and A10 Boards” on page 7-18
.
5. If the measured signal is not present, verify the receiver operation:
a. Replace the suspect receiver with a known good receiver. You can use a known good
receiver from the same analyzer.
b. If the error still occurs, replace the A35 receiver motherboard. Refer to
and Replacing the A35 Receiver Motherboard” on page 7-52
.
c. If the analyzer functions properly with the replacement receiver, replace the suspect
receiver. Refer to
“Removing and Replacing the A31, A32, A33, and A34 Receiver
.
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Страница 302: ...A 16 Service Guide E8364 90001 Error Terms PNA Series Microwave Network Analyzers Error Term Data E8362A E8363A E8364A ...
Страница 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Страница 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Страница 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...