4-28
Service Guide E8364-90001
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
E8362A, E8363A, E8364A
Verifying the A, B, R1, and R2 Traces
The first step is to verify that the A, B, R1, and R2 traces are present and that they are
approximately level:
• On the
System
menu, point to
Service
,
Utilities
, and then click
Receiver Display
.
• Traces A, B, R1, and R2 are displayed in four separate data windows as shown in
. Identifying discrepancies of the traces in these windows can help you to isolate the
faulty assembly.
Figure 4-9 Typical Four Channel Display
• If all traces are present and are similar to the traces in
, then there are no
major problems with the analyzer’s measurement system. There may, however, be a
minor failure in the analyzer.
To test further:
— Go to
Chapter 3 , “Tests and Adjustments,”
and perform all the tests in that section.
— If a problem still exists, contact Agilent. Refer to
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Страница 302: ...A 16 Service Guide E8364 90001 Error Terms PNA Series Microwave Network Analyzers Error Term Data E8362A E8363A E8364A ...
Страница 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Страница 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Страница 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...