4
Troubleshooting the E5810B
Checking the E5810B
86
E5810B User’s Guide
Checking the E5810B
Checking the E5810B front panel display
Table 4- 1
shows the suggested steps to troubleshoot the
E5810B using the front panel display.
Table 4-1
E5810B front panel display
Error message
Description
Firmware update not successful
The firmware update failed. Verify that the IP address of the FTP server,
login, user password, and filename are correct.
ERROR:NV RAM Setting defaults
Configurations in the nonvolatile data memory are corrupted. All data were
restored to the factory default settings, per the nonvolatile RAM (Random
Access Memory).
ERROR: Selftest See Syslog
A power-on self-test error occurred. Errors may be viewed in the Syslog
message by typing
http://<IP Address>/systemLog.htm
.
ERROR: No LAN connection found
LAN link integrity is not detected. Make sure that the LAN cable is
connected and the LAN connection is active.
ERROR: IP in use <
addr
>
The E5810B tried to use an IP address that is currently being used by
another device on the network.
Try one of these actions:
•
Press the Preset button and reboot the E5810B to set the E5810B back
to booting with DHCP enabled.
•
Remove other device(s) from the network and reboot the E5810B.
•
Move the E5810B to a local network, disable DHCP, and set the E5810B
to an unused IP Address.
Содержание E5810B
Страница 1: ...Agilent Technologies Agilent E5810B LAN GPIB USB Gateway User s Guide ...
Страница 8: ...VIII E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...XIV E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...XVIII E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 100: ...3 E5810B Web Access Other Web Access Functions 82 E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 156: ...Glossary V 138 E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...