Chapter 3
55
Quick Start Guide
Measurement Example of a Bandpass Filter
3. Quick S
tart
G
u
ide
STEP 2. Calibration
To turn the error correction ON, set the calibration type to the full 2-port calibration and
measure the calibration data.
For details about calibration, see Chapter “Calibration” in the
E5070B/E5071B User’s
Guide
.
Step 1.
Select the calibration kit suitable for the measurement cable. In this measurement example,
Calibration Kit 85032F is selected.
•
[Cal] - Cal Kit - 85032F
Step 2.
Set the calibration type to the full 2-port calibration using the test port 1 and 2.
•
[Cal] - Calibrate - 2-Port Cal - Select Ports - 1-2
Step 3.
Connect the OPEN standard (included in the calibration kit) to the other end of the
measurement cable that is connected to the test port 1 as shown in Figure 3-13, and
measure the open calibration data at the test port 1. After measuring the open calibration
data, symbol
√
is displayed to the left of the
Port 1 Open
button.
•
[Cal] - Calibrate - 2-Port Cal - Reflection - Port1 Open
In the same way, measure the calibration data for the SHORT/LOAD standards at the test
port 1.
Figure 3-13
Connecting the OPEN/SHORT/LOAD standards
Step 4.
In the same way as described above, measure the calibration data for the
OPEN/SHORT/LOAD standards at the test port 2.
Содержание E5070B ENA Series
Страница 8: ...8 ...
Страница 14: ...14 Chapter1 Introduction of This Manual Overview of the E5070B E5071B ...
Страница 19: ...Chapter 2 19 Installation Checking the Shipment 2 Installation Figure 2 1 E5070B E5071B Accessories ...
Страница 32: ...32 Chapter2 Installation Power Supply and Blown Fuses Figure 2 10 Power cable options ...
Страница 62: ...62 Chapter3 Quick Start Guide Measurement Example of a Bandpass Filter ...