20
Chapter 2
Installation
Environmental Requirements
Environmental Requirements
Set up the E5070B/E5071B where the following environmental requirements are met.
Operating Environments
Ensure that the operating environment meets the following requirements.
NOTE
Above environmental requirements are NOT for the specifications and measurement
accuracy of the analyzer, but for the operating environment of the analyzer.
Temperature
5
°
C to 40
°
C
Temperature range at the
error-correction
23
°
C
±
5
°
C (< 1
°
C deviation from the temperature
when performing the error-correction)
Humidity
20% to 80% at wet bulb temperature < +29
°
C
(non-condensation)
Altitude
0 to 2,000 m (0 to 6,561 feet)
Vibration
0.5 G maximum, 5 Hz to 500 Hz