background image

6

Removing Option UN3 or UN4

Use the following procedure to remove Option UN3 or UN4 as a firmware selection and to remove the 
hardware associated with Option UN3/UN4.

Removing Option UN3/UN4 in Firmware

1. Preset the signal generator

2. In the MENUS key group, press the hardkey 

Utility

.

3. In the 

Utility

 menu, press the softkey 

Instrument Adjustments

, and then select 

Hardware Options

.

The signal generator displays a list of hardware options, and an X indicates a selected option.

4. Highlight Option UN3 or UN4, then press the softkey 

Deselect Item

.

This removes the X, indicating that Option UN3/UN4 is no longer selected.

5. To disable Option UN3/UN4, press the softkey 

Proceed with Reconfiguration

.

6. To verify that you want to reconfigure the signal generator without Option UN3/UN4, press the softkey 

Confirm Change (Instrument will Reboot)

.

The signal generator reboots to the preset condition, and removes Option UN3/UN4.

Removing Option UN3/UN4 Hardware

1. Turn off the line power switch on the signal generator and disconnect the line cord.

2. Loosen the two screws on each handle and remove the two strap handles (

Figure 1 on page 7

, items 1 

and 2).

3. Remove the four bottom feet (

Figure 1

, item 3).

4. Remove the four rear feet (

Figure 1

, items 4 and 5).

5. Slide the top cover off the signal generator (

Figure 1

, item 6).

6. Remove the 11 top cover screws and remove the top cover.

7. Disconnect cables W15, W16, W17, and W19 from the A7 board (

Figure 2 on page 7

).

NOTE

If Option UND is installed, refer to 

Figure 3 on page 8

 and disconnect W15, W16, W17, W18, 

and W14 from the dual arb board (A5).

If Option UN7 is installed, refer to 

Figure 4 on page 8

 and disconnect A18W1, W21, and W23 

from the bit error rate test board (A6).

8. Remove the baseband generator board (A7).

9. Disconnect cables W3, W4, W5, and W19 from the A8 board (

Figure 5 on page 9

).

10. Disconnect W18, the rear panel interface ribbon cable, from both the data generator board (A8) and the 

rear panel interface board (A17). Discard this older version of W18.

11. Remove the data generator board (A8).

Содержание E4400-60767

Страница 1: ...0 90663 Printed in USA June 2008 Supersedes None Agilent Technologies ESG D Series Signal Generators Data Generator Assembly Replacement Kit Option UN3 UN4 to UN8 UN9 Kit Part Number E4400 60767 Model...

Страница 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or fo...

Страница 3: ...tor with 8M of RAM The installation process consists of the following basic steps 1 verifying the functionality of the signal generator 2 removing Option UN3 or UN4 3 installing Option UN8 and UN9 4 i...

Страница 4: ...assemblies should be performed at a static safe workstation Item Quantity Description Part Number Std No Opts UN9 to UN7 UN9 to UND 1 1 1 1 Board Assembly Real time I Q Baseband Generator A7 E4400 600...

Страница 5: ...Utility Error Info The first error message in the queue is shown in the text area of the display Refer to the service guide for information about the error message If there is more than one error mess...

Страница 6: ...UN3 UN4 Removing Option UN3 UN4 Hardware 1 Turn off the line power switch on the signal generator and disconnect the line cord 2 Loosen the two screws on each handle and remove the two strap handles...

Страница 7: ...7 Figure 1 Cover Removal Figure 2 Option UN3 UN4 A7 Baseband Generator Board and Cables...

Страница 8: ...8 Figure 3 Option UND A5 Dual Arb Board and Cables Figure 4 Option UN7 A6 Bit Error Rate Test Board and Cables...

Страница 9: ...9 Figure 5 Option UN3 UN4 A8 Data Generation Board and Cables...

Страница 10: ...t panel to A14P6 on the motherboard 5 Connect W5 Symbol Sync from the front panel to A14P7 on the motherboard 6 Insert the W18 ribbon cable provided in this upgrade kit between the A17 ribbon cable in...

Страница 11: ...5 INT Q cable between the motherboard connector A14P103 and the A15 daughterboard and replace it with the W25 cable provided in this upgrade kit Figure 7 on page 12 NOTE A14P103 and A14P102 are locate...

Страница 12: ...12 Figure 6 Option UN8 UN9 Top View Figure 7 Cables W25 and W26...

Страница 13: ...anel to A7P404 I OUT W17 09 Rear Panel to A7P405 DATA W3 5 Front Panel to A14P5 SYMBOL SYNC W5 6 Front Panel to A14P7 DATA CLOCK W4 7 Front Panel to A14P6 Rear Panel Interface Cable W18 Rear Panel Int...

Страница 14: ...14 Figure 8 Options UN9 UN7 UND Top View...

Страница 15: ...ted column for both Option UN8 and UN9 6 To enable the selected options press Proceed With Reconfiguration then press Confirm Change Instrument will Reboot CAUTION If you enable an option without the...

Страница 16: ...gure 1 item 6 4 Replace the four rear feet and torque the four screws to 21 in lb Figure 1 items 4 and 5 5 Replace the two strap handles one on each side of the signal generator and torque the two scr...