background image

10

Installing Option UN8 + UN9 Hardware and Cables

Refer to 

Figure 6

 and 

Figure 7 on page 12

 to assist you in installing the new A7 and A8 assemblies.

NOTE

If the signal generator has Option UND installed, you will also need to refer to 

Figure 3 on 

page 8

.

1. Insert the Board Assy-Generator/1 Meg (A8) into the A14J4 motherboard connector.

2. Insert the Real-Time I/Q Baseband Generator Board (A7) into the A14J3 motherboard connector.

3. Connect W3 (

Data

) from the front panel to A14P5 on the motherboard.

NOTE

If longer cables are required for W3, W4, and W5, use those provided in this upgrade kit.

Using a 9/16-in. socket, remove the shorter cables from the front panel. Resecure them with 
the wavey washer and hex nuts. Torque to 21 in-lb.

4. Connect W4 (

Data Clock

) from the front panel to A14P6 on the motherboard.

5. Connect W5 (

Symbol Sync

) from the front panel to A14P7 on the motherboard.

6. Insert the W18 ribbon cable provided in this upgrade kit between the A17 ribbon cable interconnect and 

A8P2.

NOTE

If Option UND is installed, reconnect W18 to both A8P2 and A5P1.

7. Connect W17 (

I Out

) to the rear panel to A7P405.

NOTE

If Option UND is installed, reconnect W17 to A5J1 instead of A7P405.

8. Connect W16 (

Q Out

) to the rear panel to A7P404.

NOTE

If Option UND is installed, reconnect W16 to A5J2 instead of A7P404.

9. Connect W15 (

BASEBAND GENREF IN

) to the rear panel to A7P403.

NOTE

If Option UND is installed, reconnect W15 to A5J3 instead of A7P403. Then connect W24 
from A5J4 to A7P403.

10. If Option UN7, bit error rate test board (A6) is installed, refer to 

Figure 4 on page 8

 and reconnect the 

following cables:

a. W23 (

BER DATA IN

) from the rear panel to A6P1

b. W22 (

BER CLK IN

) from the rear panel to A6P2

c. W21 (

BER GATE IN

) from the rear panel to A6P3

Содержание E4400-60767

Страница 1: ...0 90663 Printed in USA June 2008 Supersedes None Agilent Technologies ESG D Series Signal Generators Data Generator Assembly Replacement Kit Option UN3 UN4 to UN8 UN9 Kit Part Number E4400 60767 Model...

Страница 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or fo...

Страница 3: ...tor with 8M of RAM The installation process consists of the following basic steps 1 verifying the functionality of the signal generator 2 removing Option UN3 or UN4 3 installing Option UN8 and UN9 4 i...

Страница 4: ...assemblies should be performed at a static safe workstation Item Quantity Description Part Number Std No Opts UN9 to UN7 UN9 to UND 1 1 1 1 Board Assembly Real time I Q Baseband Generator A7 E4400 600...

Страница 5: ...Utility Error Info The first error message in the queue is shown in the text area of the display Refer to the service guide for information about the error message If there is more than one error mess...

Страница 6: ...UN3 UN4 Removing Option UN3 UN4 Hardware 1 Turn off the line power switch on the signal generator and disconnect the line cord 2 Loosen the two screws on each handle and remove the two strap handles...

Страница 7: ...7 Figure 1 Cover Removal Figure 2 Option UN3 UN4 A7 Baseband Generator Board and Cables...

Страница 8: ...8 Figure 3 Option UND A5 Dual Arb Board and Cables Figure 4 Option UN7 A6 Bit Error Rate Test Board and Cables...

Страница 9: ...9 Figure 5 Option UN3 UN4 A8 Data Generation Board and Cables...

Страница 10: ...t panel to A14P6 on the motherboard 5 Connect W5 Symbol Sync from the front panel to A14P7 on the motherboard 6 Insert the W18 ribbon cable provided in this upgrade kit between the A17 ribbon cable in...

Страница 11: ...5 INT Q cable between the motherboard connector A14P103 and the A15 daughterboard and replace it with the W25 cable provided in this upgrade kit Figure 7 on page 12 NOTE A14P103 and A14P102 are locate...

Страница 12: ...12 Figure 6 Option UN8 UN9 Top View Figure 7 Cables W25 and W26...

Страница 13: ...anel to A7P404 I OUT W17 09 Rear Panel to A7P405 DATA W3 5 Front Panel to A14P5 SYMBOL SYNC W5 6 Front Panel to A14P7 DATA CLOCK W4 7 Front Panel to A14P6 Rear Panel Interface Cable W18 Rear Panel Int...

Страница 14: ...14 Figure 8 Options UN9 UN7 UND Top View...

Страница 15: ...ted column for both Option UN8 and UN9 6 To enable the selected options press Proceed With Reconfiguration then press Confirm Change Instrument will Reboot CAUTION If you enable an option without the...

Страница 16: ...gure 1 item 6 4 Replace the four rear feet and torque the four screws to 21 in lb Figure 1 items 4 and 5 5 Replace the two strap handles one on each side of the signal generator and torque the two scr...

Отзывы: