![Agilent Technologies E1468A Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/agilent-technologies/e1468a/e1468a_user-manual_2867811080.webp)
80 Relay Life
Appendix D
•
Contact Resistance Maximum Value.
As the relay begins to wear
out, its contact resistance increases. When the resistance
exceeds a predetermined value, the relay should be replaced.
•
Contact Resistance Variance.
The stability of the contact resistance
decreases with age. Using this method, the contact resistance is
measured several (5-10) times, and the variance of the
measurements is determined. An increase in the variance
indicates deteriorating performance.
•
Number of Relay Operations.
Relays can be replaced after a
predetermined number of contact closures. However, this method
requires knowledge of the applied load and life specifications for
the applied load.