Executing Several Waveform Segments
The MULSEG program shows how to generate an arbitrary waveform using
two different waveform segments. One waveform segment generates a 1 kHz,
5 V sine wave repeated twice. The other one generates a 1 kHz, 5 Vpeak
triangle repeated once. The commands are:
1. Reset the AFG
*RST
The
*RST
command aborts waveform output and sets the AFG to a
defined state.
2. Clear the AFG Memory of All Sequence Data
[SOURce:]LIST[1]:SSEQuence:DELete:ALL
This command clears all segment sequence data stored in the
sequence memory.
3. Clear the AFG Memory of All Segment Data
[SOURce:]LIST[1]:DELete:ALL
This command clears all segment data stored in the segment memory.
4. Set the Sample Rate
[SOURce:]FREQuency[1][:CW | :FIXed] <
frequency
>
This command sets the rate at which the points are output by the
AFG. The frequency is:
(sample frequency) / (number of points)
Refer to Table B-3 in Appendix B for the frequency limits.
Chapter 3
Generating Arbitrary Waveforms 93
Содержание E1445A
Страница 2: ......
Страница 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Страница 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Страница 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 54: ...Notes 52 Getting Started Chapter 1...
Страница 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Страница 204: ...Notes 202 Arming and Triggering Chapter 5...
Страница 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Страница 446: ...Notes 444 AFG Status Chapter 9...
Страница 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Страница 534: ...Notes 532 Agilent E1445A User s Manual Index...