Available Marker Sources
There are six marker sources available for output at the AFG’s front panel
“Marker Out” BNC and the ECL Trigger Lines. Use the
[SOURce:]MARKer:FEED <
source
>
command to select the “Marker Out”
BNC; use
[SOURce:]MARKer:ECLTrg<
n
>:FEED <
source
>
command to
select the ECL trigger lines.
The sources for the “Marker Out” BNC, not the ECL trigger lines, can
output the marker pulses as either active high (
NORMal
) or active low
(
INVerted
). Use the
[SOURce:]MARKer:POLarity <
polarity
>
command to
select the polarity.
The different marker sources are as follows:
ARM[:STARt|:SEQuence[1]][:LAYer[1]]
For arbitrary waveforms, the marker level changes with the first point on the
waveform of the first waveform repetition. The source then outputs a marker
pulse at the last waveform point of each repetition. For
SINUsoid
outputs, the
marker is a 50% duty cycle square wave at the waveform frequency.
ARM[:STARt|:SEQuence[1]]:LAYer2
The AFG asserts a marker when triggering the first amplitude point after
receiving a start arm. The AFG unasserts a marker with the last amplitude
point of the last waveform repetition, or following an
ABORt
.
[SOURce:]FREQuency[1]:CHANge
The source outputs a one sample period wide marker pulse after a frequency
change occurs. This shows that the steady state frequency was reached.
[SOURce:]LIST[1]
The source outputs marker pulses specified by the
[SOURce:]LIST[1][:SEGMent]:MARKer
and
[SOURce:]LIST[1]:SSEQuence:MARKer
commands. Increase the pulse size
by selecting marker output for consecutive points on the waveform. Can
only be used with arbitrary waveforms (see Chapters 3 and 7 on how to
generate arbitrary waveforms).
[SOURce:]PM:DEViation:CHANge
This source outputs a one sample period wide marker pulse after a phase
change occurs. This shows that the new phase was reached.
Chapter 6
Marker Outputs/Multiple AFG Operations 205
Содержание E1445A
Страница 2: ......
Страница 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Страница 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Страница 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 54: ...Notes 52 Getting Started Chapter 1...
Страница 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Страница 204: ...Notes 202 Arming and Triggering Chapter 5...
Страница 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Страница 446: ...Notes 444 AFG Status Chapter 9...
Страница 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Страница 534: ...Notes 532 Agilent E1445A User s Manual Index...