PTRansition
sets the positive transition. For each bit unmasked, a 0 to 1
transition of that bit in the Condition Register sets the associated bit in the
Event Register.
<
unmask
>
is the decimal, hexadecimal (#H), octal (#Q), or binary (#B)
value of the Condition Register bit to be unmasked. (Bits 0, 3, 6, and 8 have
corresponding decimal values of 1, 8, 64, and 256.)
The Event Register
The Event Register latches transition events from the Condition Register as
specified by the Transition Filter. Bits in the Event Register are latched and
remain set until the register is cleared by one of the following commands:
STATus:OPERation[:EVENt]?
*CLS
The Enable Register
The Enable Register specifies which bits in the Event Register can generate
a summary bit which is subsequently used to generate a service request.
The AFG logically ANDs the bits in the Event Register with bits in the
Enable Register, and ORs the results to obtain a summary bit.
The bits in the Enable Register that are to be ANDed with bits in the Event
Register are specified (unmasked) with the command:
STATus:OPERation:ENABle <
unmask
>
<
unmask
>
is the decimal, hexadecimal (#H), octal (#Q), or binary (#B)
value of the Enable Register bit to be unmasked. (Bits 0, 3, 6, and 8 have
corresponding decimal values of 1, 8, 64, and 256.)
The Enable Register is cleared at power-on, or by specifying an
<
unmask
>
value of 0.
Program Example
The OSG_RQS program sets up the Operation Status Group Registers to
determine when the AFG enters the wait-for-arm state. When the AFG
enters that state, a service request interrupt is sent to the computer which
responds with a message indicating the state which exists.
The steps of the program are:
1. Set the bit transition which will latch the event (entering wait-for-arm
state) in the Event Register.
STATus:OPERation:NTRansition <
unmask
>
or
STATus:OPERation:PTRansition <
unmask
>
436 AFG Status
Chapter 9
Содержание E1445A
Страница 2: ......
Страница 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Страница 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Страница 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Страница 54: ...Notes 52 Getting Started Chapter 1...
Страница 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Страница 204: ...Notes 202 Arming and Triggering Chapter 5...
Страница 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Страница 446: ...Notes 444 AFG Status Chapter 9...
Страница 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Страница 534: ...Notes 532 Agilent E1445A User s Manual Index...