Repair and Maintenance Guidelines
This section provides guidelines for handling the relay multiplexers
including:
•
ESD precautions
•
Soldering printed circuit boards
•
Post-repair safety checks
ESD Precautions
Electrostatic discharge (ESD) may damage static sensitive devices in the
multiplexers. This damage can range from slight parameter degradation to
catastrophic failure. When handling multiplexer assemblies, observe the
following guidelines to avoid damaging multiplexer components:
•
Always use a static-free work station with a pad of conductive rubber
or similar material when handling multiplexer components.
•
If a device requires soldering, be sure the assembly is placed on a
pad of conductive material. Also, be sure that you, the pad, and the
soldering iron tip are grounded to the assembly.
Soldering Printed
Circuit Boards
Some of the components on the etched circuit boards of the multiplexer
modules have plated through-holes that provide a solder path to both sides
of the insulating material. Soldering can be done from either side of the
board with equally good results. When soldering to any circuit board, keep
in mind the following guidelines:
•
Avoid unnecessary component unsoldering and soldering. Excessive
replacement can result in damage to the circuit board, adjacent
components, or both.
•
Do not use a high power soldering iron on etched circuit boards, as
excessive heat may lift a conductor or damage the board.
•
Use a suction device or wooden toothpick to remove solder from
component mounting holes. When using a suction device, be sure
that the equipment is properly grounded.
Post-Repair Safety
Checks
After making repairs to the multiplexer, inspect the multiplexer for any
signs of abnormal internally generated heat, such as discolored printed
circuit boards or components, damaged insulation, or evidence of arcing.
Determine and correct the cause of the condition. Then perform the
functional tests as described in Chapter 2, "Verification Tests," to verify that
the multiplexer is functional.
Chapter 4
Service 59
Содержание Agilent E1345A
Страница 2: ......
Страница 8: ...Notes 6 E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 9: ......
Страница 10: ......
Страница 12: ...Notes 10 What s In This Manual E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 20: ...Figure 1 2 Initial Incoming Inspection Guidelines 18 General Information Chapter 1 ...
Страница 22: ...Notes 20 General Information Chapter 1 ...
Страница 49: ...Figure 3 2 E1343A E1344A E1345A E1347A Terminal Module Mechanical Parts Chapter 3 Replaceable Parts 47 ...
Страница 50: ...Figure 3 3 E1355A E1356A Terminal Module Mechanical Parts 48 Replaceable Parts Chapter 3 ...
Страница 54: ...Figure 4 1 E1343A E1344A E1345A E1347A Multiplexer Block Diagram 52 Service Chapter 4 ...
Страница 56: ...Figure 4 2 E1355A 56A Multiplexers Block Diagram 54 Service Chapter 4 ...
Страница 64: ...Notes 62 Service Chapter 4 ...
Страница 72: ...Notes 70 Example C Programs Appendix A ...
Страница 81: ...Figure B 2 E1343A 44A 45A 47A Terminal Module Mechanical Parts Appendix B Backdating Information 79 ...
Страница 82: ...Figure B 3 E1355A 56A Terminal Module Mechanical Parts 80 Backdating Information Appendix B ...
Страница 83: ......
Страница 84: ...Copyright Agilent Technologies Inc 2005 E1345 90013 E1345 90013 ...