Test 2-1: Closed Channel Resistance Test
This test verifies that all relay contacts meet the closed-channel resistance
specification for the multiplexer. If the closed-channel resistance of any
relay contact is greater than 2
Ω
, the relay should be replaced. This test uses
the test fixture (see Figure 2-1).
Measuring Protection
Resistors
Since there are 100
Ω
protection resistors (R20 through R31) in the relay
paths, measure the protection resistor values to begin this test. The values of
the protection resistors are then subtracted from the measured path
resistance to determine the relay contact resistance. To measure the
protection resistor values, set the 3458A DMM to 4-wire ohms, autorange
and measure each resistor value with the DMM (see Figure 2-2). Record the
measured values in Table 2-1.
NOTE
On the E1343A and E1344A, Resistors R22, R25, R28, and R31 are 1.0k
Ω
resistors.
Table 2-1. Measured Protection Resistor Values
Resistor
Measured Value (
Ω
)
Resistor
Measured Value (
Ω
)
R20
R21
R22
___________________
___________________
___________________
R26
R27
R28
___________________
___________________
___________________
R23
R24
R25
___________________
___________________
___________________
R29
R30
R31
___________________
___________________
___________________
Figure 2-2. Typical Resistance Measurement Connections
24 Verification Tests
Chapter 2
Содержание Agilent E1345A
Страница 2: ......
Страница 8: ...Notes 6 E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 9: ......
Страница 10: ......
Страница 12: ...Notes 10 What s In This Manual E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 20: ...Figure 1 2 Initial Incoming Inspection Guidelines 18 General Information Chapter 1 ...
Страница 22: ...Notes 20 General Information Chapter 1 ...
Страница 49: ...Figure 3 2 E1343A E1344A E1345A E1347A Terminal Module Mechanical Parts Chapter 3 Replaceable Parts 47 ...
Страница 50: ...Figure 3 3 E1355A E1356A Terminal Module Mechanical Parts 48 Replaceable Parts Chapter 3 ...
Страница 54: ...Figure 4 1 E1343A E1344A E1345A E1347A Multiplexer Block Diagram 52 Service Chapter 4 ...
Страница 56: ...Figure 4 2 E1355A 56A Multiplexers Block Diagram 54 Service Chapter 4 ...
Страница 64: ...Notes 62 Service Chapter 4 ...
Страница 72: ...Notes 70 Example C Programs Appendix A ...
Страница 81: ...Figure B 2 E1343A 44A 45A 47A Terminal Module Mechanical Parts Appendix B Backdating Information 79 ...
Страница 82: ...Figure B 3 E1355A 56A Terminal Module Mechanical Parts 80 Backdating Information Appendix B ...
Страница 83: ......
Страница 84: ...Copyright Agilent Technologies Inc 2005 E1345 90013 E1345 90013 ...