Functional Verification Test
The Functional Verification Test for the multiplexer modules consists of
sending the
*IDN?
command and checking the response. This test can be
used to verify that the multiplexer is connected properly and is responding
to a basic command.
Procedure
1. Verify that the multiplexer is properly installed in mainframe
2. Verify that the mainframe has passed its power-on test.
3. Send
*IDN?
to the multiplexer (see example following)
4. The return should be as follows (revision number may vary):
HEWLETT-PACKARD,SWITCHBOX,0,A.06.00
NOTE
If the primary address setting, secondary address setting, or the interface
select code is set incorrectly, the multiplexer will not respond. Verify
proper address selection before troubleshooting.
Example
An example follows which uses an HP 9000 Series 300 computer with
HP BASIC and a multiplexer address of 70914.
10 DIM A$[100]
20 OUTPUT 70914;"*IDN?"
!Send the ID query command
30 ENTER 70914;A$
!Get response
40 PRINT A$
!Print response
50 END
Operation Verification Test
The procedures in this section are used to provide a high level of confidence
that the multiplexer is meeting published specifications. The Operation
Verification Test is a subset of the Performance Verification Tests and is
suitable for checkout after performing repairs.
The Operation Verification Test is performed by completing the Closed
Channel Resistance Test (Test 2-1) as described in the Performance
Verification Test procedures. This test is usually sufficient to verify that the
multiplexer is meeting its specifications.
22 Verification Tests
Chapter 2
Содержание Agilent E1345A
Страница 2: ......
Страница 8: ...Notes 6 E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 9: ......
Страница 10: ......
Страница 12: ...Notes 10 What s In This Manual E1343A 44A 45A 47A 55A 56A Service Manual ...
Страница 20: ...Figure 1 2 Initial Incoming Inspection Guidelines 18 General Information Chapter 1 ...
Страница 22: ...Notes 20 General Information Chapter 1 ...
Страница 49: ...Figure 3 2 E1343A E1344A E1345A E1347A Terminal Module Mechanical Parts Chapter 3 Replaceable Parts 47 ...
Страница 50: ...Figure 3 3 E1355A E1356A Terminal Module Mechanical Parts 48 Replaceable Parts Chapter 3 ...
Страница 54: ...Figure 4 1 E1343A E1344A E1345A E1347A Multiplexer Block Diagram 52 Service Chapter 4 ...
Страница 56: ...Figure 4 2 E1355A 56A Multiplexers Block Diagram 54 Service Chapter 4 ...
Страница 64: ...Notes 62 Service Chapter 4 ...
Страница 72: ...Notes 70 Example C Programs Appendix A ...
Страница 81: ...Figure B 2 E1343A 44A 45A 47A Terminal Module Mechanical Parts Appendix B Backdating Information 79 ...
Страница 82: ...Figure B 3 E1355A 56A Terminal Module Mechanical Parts 80 Backdating Information Appendix B ...
Страница 83: ......
Страница 84: ...Copyright Agilent Technologies Inc 2005 E1345 90013 E1345 90013 ...