TECHNICAL INFORMATION
For these parameters the possible settings and the
messages displayed are the same shown for the
MANUAL and AUTOMATIC operating mode,
excepted for the script
A/R
in place of
MAN
at the
end of the second row.
Parameters Values
In the following table are indicated the different
possible values for each sublimation parameter.
PARAMETER
DEFAULT VALUE
ADJUSTABLE RANGE
Current
30 A
30 A to 50 A at step of 0.5 A
Time
1 minute
1 minute to 7 minutes at step of
30 seconds
Period
3 minutes
3 minutes to 32 hours with
following intermediate values:
−
10 minutes
−
30 minutes
−
1 hour
−
2 hours
−
4 hours
−
8 hours
Pressure threshold
1x10
-7
mbar
1x10
-10
mbar to 1x10
-4
mbar at
step of 1x10
-X
mbar
Sublimation
To start the sublimation proceed as follows:
−
On the rear panel, turn the power switch on (I
position).
−
Wait for the following messages displayed in
sequence:
A
U
T
O
T E
S
T
A
U
T
O
T E
S
T
O K
M i n i T i - b a l l
v
M A N
Or:
T S P
F i l a m e n t X
v
M A N
Then press the
SUBLIMATION
key. The relevant
Led will be turned on and the display shows:
S U B L I M A T I O N
v
S T A R T ? N O M A N
Pressing the
W
or
V
keys the display shows:
S U B L I M A T I O N
v
S T A R T ? Y E S M A N
Press
SET
key to confirm the start or any other key
to escape. After the start time, during the current
ramp the display shows:
V = X X . X V I = X X . X A
C U R R .
R A M P M A N
Where
XX.X
are the currently value used.
Then, finish the current ramp, during the
sublimation, the display shows the voltage, the
current and the time used for the sublimation, as
follows:
V = X X . X V I = X X . X A
T = X X : X X : X X M A N
Where
XX.X
are the currently value used and
XX:XX:XX
is the sublimation time expressed in
hours, minutes and seconds format.
23 87-900-077-01(D)