3-12
Verifying Performance
Understanding the Tests
clock signal. The option does not have GPIB control so the program cannot verify
that the option is operational.
Test A: RF Analyzer
Pulse Demodulation
Equipment Required
HP/Agilent 8657A/B Signal Generator
HP/Agilent 8116A Pulse Generator
Theory of the Test
The Pulse generator drives the signal generator to generate Pulsed RF signals at
various frequencies. The UUT demodulates these signals and the internal
oscilloscope measures the rise and fall time of the demodulated waveform. This test
verifies the demodulator rise and fall time specifications.
Things To Check In Case Of Problems
•
The oscilloscope measurement functions are used to measure the 10% to 90% rise
and fall time. If the demodulated waveform has significant video feedthrough or
the oscilloscope cannot trigger the measurements properly, an incorrect
measurement may result.
Test B: Spectrum
Analyzer (Option
006 only)
Equipment Required
HP/Agilent 8657A/B Signal Generator
Theory of the Test
The signal generator is used to stimulate the UUT spectrum analyzer at various
frequencies. At each frequency, the spectrum analyzer is tuned to measure its own
image and spurious responses. The level of the signal generator is set to various
levels to determine the spectrum analyzer’s log-linearity.
Things To Check In Case Of Problems
•
The absolute accuracy of the signal generator causes a significant measurement
uncertainty in the log linearity measurements. A more accurate measurement is
to reduce the signal level using a precision attenuator. This is an acceptable
secondary manual test if the UUT fails the automated test.
Содержание 8922M
Страница 8: ...Contents Contents 6...
Страница 21: ...1 1 1 Installing Your Agilent 8922M S...
Страница 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Страница 29: ...2 1 2 Making Measurements...
Страница 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Страница 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Страница 75: ...2 47 Making Measurements Advanced Features...
Страница 76: ...3 1 3 Verifying Performance...
Страница 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Страница 224: ...4 118 Screens Service Service This screen is documented in the service documentation To escape press PREV...
Страница 235: ...4 129 Screens Tests Tests Instrument BASIC Refer to chapter 8 Instrument BASIC for information about the Tests Screen...
Страница 236: ...4 130 Screens Tests...
Страница 237: ...5 1 5 Keys...
Страница 238: ...5 2 Keys Key Map Key Map...
Страница 275: ...6 27 Connectors Timing Diagrams Timing Diagrams Figure 6 2 Digital Demodulation Timing Specification Table...
Страница 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Страница 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Страница 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Страница 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Страница 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Страница 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Страница 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Страница 283: ...6 35 Connectors Timing Diagrams Figure 6 10 Reset and Hold Hop Address Source Int Timing Diagram...
Страница 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Страница 297: ...8 1 8 Instrument BASIC...
Страница 341: ...A 1 A APPENDIX A...
Страница 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Страница 365: ...B 1 B Glossary...
Страница 370: ...B 6 Glossary...