4-15
Screens
Bit Error 2
24. MS Loopback
This field is used to enter or display amount of delay (expressed in number of speech
frames), before bits are compared. The Agilent 8922 uses this value to synchronize
to the correct received PRBS speech frame. The correct number of speech frames
can be determined automatically by making a Bit Error Test using Auto Mode when
the bit error ratio is approximately <20%. Then, if faster measurements are desired
or if the bit error ratio exceeds 20%, change the mode to manual.
See Also
Screens: Bit Error Test (Manual, Auto)
23. Bits Tested
(Completed)
This field displays the number of bits that were tested during the last fully completed
test.
Bit error tests may extend beyond the number entered in the “Bits To Test” field so
that the last speech frame may be completed.
Содержание 8922M
Страница 8: ...Contents Contents 6...
Страница 21: ...1 1 1 Installing Your Agilent 8922M S...
Страница 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Страница 29: ...2 1 2 Making Measurements...
Страница 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Страница 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Страница 75: ...2 47 Making Measurements Advanced Features...
Страница 76: ...3 1 3 Verifying Performance...
Страница 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Страница 224: ...4 118 Screens Service Service This screen is documented in the service documentation To escape press PREV...
Страница 235: ...4 129 Screens Tests Tests Instrument BASIC Refer to chapter 8 Instrument BASIC for information about the Tests Screen...
Страница 236: ...4 130 Screens Tests...
Страница 237: ...5 1 5 Keys...
Страница 238: ...5 2 Keys Key Map Key Map...
Страница 275: ...6 27 Connectors Timing Diagrams Timing Diagrams Figure 6 2 Digital Demodulation Timing Specification Table...
Страница 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Страница 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Страница 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Страница 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Страница 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Страница 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Страница 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Страница 283: ...6 35 Connectors Timing Diagrams Figure 6 10 Reset and Hold Hop Address Source Int Timing Diagram...
Страница 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Страница 297: ...8 1 8 Instrument BASIC...
Страница 341: ...A 1 A APPENDIX A...
Страница 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Страница 365: ...B 1 B Glossary...
Страница 370: ...B 6 Glossary...