6
-
2
Performance Tests
Performance Tests
The procedures in this chapter test the performance of
the Agilent 86121A using the specifications listed in
Chapter 7, “Specifications and Regulatory Information”
as the performance standard. All of the tests are done
manually without the aid of a computer. None of these
tests require access to the interior of the instrument.
Test 1. Absolute Wavelength Accuracy
Test 3. Polarization Dependence
Test 4. Optical Input Return Loss
6-8
Test 5. Amplitude Accuracy and Linearity
6-11
Allow the 86121A to warm up for 15 minutes before
doing any of the performance tests.
Calibration Cycle
This instrument requires periodic verification of perfor-
mance. The instrument should have a complete verifi-
cation of specifications at least once every year.
Connector Cleaning
Accurate and repeatable measurements require clean
connections. Follow the connector cleaning procedure,
refer to “Cleaning Connections for Accurate Measure-
ments” on page 8-11
, before proceeding with the per-
formance verification.
Содержание 86121A
Страница 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Страница 7: ......
Страница 10: ...Chapter 1 Getting Started ...
Страница 21: ......
Страница 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Страница 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Страница 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Страница 32: ...The Softkeys 2 11 The Setup menu ...
Страница 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Страница 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Страница 37: ......
Страница 191: ......
Страница 205: ...7 14 Regulatory Information Declaration of Conformity ...
Страница 239: ......