Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
91
Broadband Component
(Isolator, Circulator, Coupler, Attenuator and so on)
Characteristics
• Wide wavelength measurement range
• Usually low CD and PMD, but especially PMD may need to be
verified
• Loss may be high, but relatively independent of wavelength
Measurement Tips
The broadband nature of these devices allows you to trade off
wavelength resolution for faster measurement speed.
• Set the modulation frequency in manual mode to 2000 MHz.
• In either swept or stepped mode, optimize measurement
speed by setting the wavelength increment just small enough
to resolve the wavelength dependencies of the attributes
being measured.
• If you use the smoothing function to reduce trace noise, you
may need to increase the number of trace points. (for
example, reduce the wavelength increment) to avoid loss of
resolution.
Содержание 86038B
Страница 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Страница 4: ...4 ...
Страница 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Страница 58: ...58 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 11 86038B Block Diagram ...
Страница 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Страница 230: ...230 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition VB 6 0 Syntax pdlaClient Actions PrintScreen ...
Страница 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 349: ......