![Agilent Technologies 85225F Скачать руководство пользователя страница 143](http://html.mh-extra.com/html/agilent-technologies/85225f/85225f_installation-and-user-manual_2867939143.webp)
Network Analyzer Performance Specification Summary
G
Installation and User’s Guide
143
Measurement Port Characteristics
Measurement Uncertainty
Measurement uncertainty curves utilize an RSS (Root Sum Square) model
for the contribution of random errors such as noise and typical connector
and test set switch repeatabilities. These are combined with a worst- case
model for the contributions of dynamic accuracy and residual systemic
errors.
Curves show the worst- case magnitude and phase uncertainty for
reflection and transmission measurements, using the specified cal kit, with
10 Hz IF bandwidth, and no averaging during the measurement.
Reflection Measurements
Frequency Range (GHz)
Residual
0.045 to 2.0
2.0 to 20.0
20.0 to 40.0
40.0 to 50.0
Directivity
42 dB
42 dB
38 dB
36 dB
Source match
41 dB
38 dB
33 dB
31 dB
Load match
42 dB
42 dB
37 dB
35 dB
Reflection tracking
±(0.001 + 0.2 dB/
°
C)
±(0.008 + 0.2 dB/
°
C)
±(0.020 + 0.3 dB/
°
C)
±(0.027 + 0.4 dB/
°
C)
Transmission tracking
±(0.019 + 0.2 dB/
°
C)
±(0.053 + 0.2 dB/
°
C)
±(0.114 + 0.3 dB/
°
C)
±(0.215 + 0.4 dB/
°
C)
Содержание 85225F
Страница 1: ...Agilent Technologies Agilent 85225F PerformanceModelingSystem Installation and User s Guide ...
Страница 90: ...90 Installation and User s Guide 2 Installing the System ...
Страница 102: ...102 Installation and User s Guide 3 Verifying System Functionality ...
Страница 110: ...110 Installation and User s Guide 4 Servicing the System ...
Страница 118: ...118 Installation and User s Guide A Enhancing Measurement Accuracy ...
Страница 128: ...128 Installation and User s Guide C RF Subsystem Functional Verification Test ...
Страница 144: ...144 Installation and User s Guide G Network Analyzer Performance Specification Summary Transmission Measurements ...
Страница 148: ...148 Installation and User s Guide Index ...