85130C Operating and Service Manual

Equipment and Supplies

The following equipment and supplies are required for the maintenance and use of, but are not supplied 
with, your 85130C adapter. kit.

Contacting Agilent

Assistance with test and measurement needs and information on finding a local Agilent office are available 
on the Web at:

www.agilent.com/find/assist

NOTE

In any correspondence or telephone conversation, refer to the Agilent product by its model 
number and full serial number. With this information, the Agilent representative can 
determine whether your product is still within its warranty period.

Item

Part Number

3.5 mm gage kit (part of the 85052B calibration kits)

85052-80010

Type-N gage kit (part of the 85054B calibration kit)

85054-80011

Torque wrench, 3/4”, 136 N-cm (12 in-lb) (part of the 85054B calibration kit)

8710-1766

Torque wrench, 20 mm, 96 N-cm (8 in-lb)

8710-1764

Document:

 Connector Care for RF and Microwave Coaxial Connectors

08510-90064

Содержание 85130C

Страница 1: ...Agilent Technologies Agilent 85130C NMD 3 5 mm to Type N Adapter Kit Operating and Service Manual...

Страница 2: ...ument are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government p...

Страница 3: ...o the Connector Care for RF and Microwave Coaxial Connectors document It can be viewed online by searching for part number 08510 90064 at www agilent com Description The 85130C adapters are used when...

Страница 4: ...e male connector See Figure 1 on page 3 Zero the gages before each use by attaching the appropriate gage master to the end of the gage torquing the connection to 12 in lb with a 3 4 inch torque wrench...

Страница 5: ...dically repeat the test to determine if the performance meets the electrical specifications on the previous page or if they need to be replaced An initial period of one year between performance tests...

Страница 6: ...functions mentioned in the return loss test below 1 Preset the analyzer 2 Set a stimulus start frequency of the analyzer s lowest frequency 3 Set a stimulus stop frequency of 18 GHz 4 Set an IF bandw...

Страница 7: ...laceable Parts There are no replaceable components for the adapters A worn or damaged adapter must be replaced in whole ELECTRICAL SPECIFICATIONS Tested by ______________ Date ______________ Adapter F...

Страница 8: ...espondence or telephone conversation refer to the Agilent product by its model number and full serial number With this information the Agilent representative can determine whether your product is stil...