Agilent Technologies 85130C Скачать руководство пользователя страница 4

 

2

85130C Operating and Service Manual

Specifications

Agilent Technologies guarantees that your adapters will equal or exceed the following specifications in the 
+20

 to +26

C (+68

 to +79

F) temperature range.

The allowable recession of the center conductor of the NMD-3.5 mm (f) connector is –0.0006 to –0.0022 
inches (–0.015 to –0.056 mm) below the mating surface of the outer conductor.

The allowable pin position for type-N connectors is:

male = –0.2075 to –0.2079 inch (–5.2705 to –5.28066 mm)

female = +0.2061 to +0.2065 inch (+5.23494 to +5.2451 mm)

In a mated pair of type-N connectors, the center conductor’s mating plane is offset from the outer 
conductor’s mating plane by 0.207 inch (5.2578 mm) in the direction of the male connector. See 

Figure 1 on 

page 3

.

Zero the gages before each use by attaching the appropriate gage master to the end of the gage, torquing 
the connection to 12 in-lb with a 3/4 inch torque wrench, and adjusting the dial on the face of the gage so 
the gage reads zero.

Both type-N gages, male and female, read zero when the center conductor to outer conductor offset is 
nominally 0.2070 inch (5.2578 mm).

When gaging a male connector, the actual recession of the center conductor is –0.2070 inch 

plus

 the 

reading on the gage.

When gaging a female connector, the actual recession of the center conductor is +0.2070 inch 

plus

 the 

reading on the gage.

Example: for a gage reading of –0.0001 inch, the following is true:

Frequency

Return Loss

DC – 8 GHz

 

34 dB

8 – 18 GHz

 

28 dB

Gage Reading

Actual Pin Position

(versus outer mating plane)

--

Male

Female

–0.0001 inch

–0.2070 + 
(–0.0001) = 
–0.2071 inch

+0.2070 + 
(–0.0001) = 
+0.2069 inch

Содержание 85130C

Страница 1: ...Agilent Technologies Agilent 85130C NMD 3 5 mm to Type N Adapter Kit Operating and Service Manual...

Страница 2: ...ument are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government p...

Страница 3: ...o the Connector Care for RF and Microwave Coaxial Connectors document It can be viewed online by searching for part number 08510 90064 at www agilent com Description The 85130C adapters are used when...

Страница 4: ...e male connector See Figure 1 on page 3 Zero the gages before each use by attaching the appropriate gage master to the end of the gage torquing the connection to 12 in lb with a 3 4 inch torque wrench...

Страница 5: ...dically repeat the test to determine if the performance meets the electrical specifications on the previous page or if they need to be replaced An initial period of one year between performance tests...

Страница 6: ...functions mentioned in the return loss test below 1 Preset the analyzer 2 Set a stimulus start frequency of the analyzer s lowest frequency 3 Set a stimulus stop frequency of 18 GHz 4 Set an IF bandw...

Страница 7: ...laceable Parts There are no replaceable components for the adapters A worn or damaged adapter must be replaced in whole ELECTRICAL SPECIFICATIONS Tested by ______________ Date ______________ Adapter F...

Страница 8: ...espondence or telephone conversation refer to the Agilent product by its model number and full serial number With this information the Agilent representative can determine whether your product is stil...

Отзывы: