85130C Operating and Service Manual

 3

Figure 1 

Type-N Connectors

Proper Use

Attach the adapters to the test ports and tighten them finger tight. Use the spanner wrench to hold the test 
set end of the adapter and torque the test set connector with a 20 mm torque wrench set to 96 N-cm (8 
in-lb). 

Performance Tests

Use a network analyzer to perform the following return loss test on your adapters as soon as you receive 
them. Periodically repeat the test to determine if the performance meets the electrical specifications on the 
previous page, or if they need to be replaced. An initial period of one year between performance tests is 
recommended.

Required Equipment

Agilent Model or Part Number

Network analyzer – 

18 GHz (or higher) 
measurement capability – with 
time domain option 010

PNA with Option 010 (See 

PNA 

Family Microwave Network 
Analyzers Configuration Guide

part number 5990-7745EN)

Type-N loads (included in the 
85054B calibration kit)

00909-60011 (male)

00909-60012 (female)

Type-N 50 ohm airline, 12.5 cm 
(included in the 85055A 
verification kit)

85055-60001

Содержание 85130C

Страница 1: ...Agilent Technologies Agilent 85130C NMD 3 5 mm to Type N Adapter Kit Operating and Service Manual...

Страница 2: ...ument are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government p...

Страница 3: ...o the Connector Care for RF and Microwave Coaxial Connectors document It can be viewed online by searching for part number 08510 90064 at www agilent com Description The 85130C adapters are used when...

Страница 4: ...e male connector See Figure 1 on page 3 Zero the gages before each use by attaching the appropriate gage master to the end of the gage torquing the connection to 12 in lb with a 3 4 inch torque wrench...

Страница 5: ...dically repeat the test to determine if the performance meets the electrical specifications on the previous page or if they need to be replaced An initial period of one year between performance tests...

Страница 6: ...functions mentioned in the return loss test below 1 Preset the analyzer 2 Set a stimulus start frequency of the analyzer s lowest frequency 3 Set a stimulus stop frequency of 18 GHz 4 Set an IF bandw...

Страница 7: ...laceable Parts There are no replaceable components for the adapters A worn or damaged adapter must be replaced in whole ELECTRICAL SPECIFICATIONS Tested by ______________ Date ______________ Adapter F...

Страница 8: ...espondence or telephone conversation refer to the Agilent product by its model number and full serial number With this information the Agilent representative can determine whether your product is stil...