Agilent Technologies 85130C Скачать руководство пользователя страница 6

 

4

85130C Operating and Service Manual

Figure 2 

Return Loss Setup

Return loss is measured by connecting a 50 ohm fixed load termination through a 12.5 cm airline to the 
adapter, then attaching the adapter to port one of the test set (see 

Figure 2

).

The effects of an imperfect load may be gated out using the network analyzer time domain option.

NOTE

Refer to your network analyzer’s Help system for specific instructions on using the functions 
mentioned in the return loss test below.

1. Preset the analyzer.

2. Set a stimulus start frequency of the analyzer’s lowest frequency.

3. Set a stimulus stop frequency of 18 GHz.

4. Set an IF bandwidth of 100 Hz.

5. Perform, and then save, a 3.5 mm 1-port S

11

 calibration.

6. With correction turned on, select the time domain mode.

7. Set a stimulus start time for the sweep to –0.05 nano-seconds.

8. Set a stimulus stop time for the sweep to 1.9 nano-seconds.

9. Select the gating function and gate-out everything but the adapter. See 

Figure 3

.

10. Select the analyzer’s frequency domain mode.

11. Use the markers to read the return loss value.

Содержание 85130C

Страница 1: ...Agilent Technologies Agilent 85130C NMD 3 5 mm to Type N Adapter Kit Operating and Service Manual...

Страница 2: ...ument are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government p...

Страница 3: ...o the Connector Care for RF and Microwave Coaxial Connectors document It can be viewed online by searching for part number 08510 90064 at www agilent com Description The 85130C adapters are used when...

Страница 4: ...e male connector See Figure 1 on page 3 Zero the gages before each use by attaching the appropriate gage master to the end of the gage torquing the connection to 12 in lb with a 3 4 inch torque wrench...

Страница 5: ...dically repeat the test to determine if the performance meets the electrical specifications on the previous page or if they need to be replaced An initial period of one year between performance tests...

Страница 6: ...functions mentioned in the return loss test below 1 Preset the analyzer 2 Set a stimulus start frequency of the analyzer s lowest frequency 3 Set a stimulus stop frequency of 18 GHz 4 Set an IF bandw...

Страница 7: ...laceable Parts There are no replaceable components for the adapters A worn or damaged adapter must be replaced in whole ELECTRICAL SPECIFICATIONS Tested by ______________ Date ______________ Adapter F...

Страница 8: ...espondence or telephone conversation refer to the Agilent product by its model number and full serial number With this information the Agilent representative can determine whether your product is stil...

Отзывы: