![Agilent Technologies 85024A Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925029.webp)
Chapter 4
4-5
Operation
Operator’s Check
Operator’s Check Using a Spectrum Analyzer
Required Equipment
Procedure
1. Connect the spectrum analyzer’s calibration output directly to its RF input. Turn on the spectrum
analyzer.
2. Set the spectrum analyzer center frequency to the frequency of the calibration output.
3. Set the frequency span to 10 MHz and activate a marker.
4. Place the marker on the peak of the displayed signal and note the value in dBm.
5. Connect the probe adapter to the calibration output.
6. Connect the probe output to the RF input on the analyzer.
7. Connect the probe power cable to a power source.
8. Plug the probe tip into the probe adaptor.
9. Place the marker on the peak of the displayed signal and compare the signal level in dBm to that
measured in step 4. The probe should not cause more than 5 dBm of signal loss.
Item
Part Number
Spectrum Analyzer
Any Compatible
Adapters
As Necessary
Probe Adapters
11880-60001
Содержание 85024A
Страница 4: ...iv ...
Страница 7: ...1 1 1 General Information ...
Страница 13: ...2 1 2 Accessories ...
Страница 19: ...3 1 3 Installation ...
Страница 24: ...3 6 Chapter3 Installation Returning the Product for Service ...
Страница 25: ...4 1 4 Operation ...
Страница 30: ...4 6 Chapter4 Operation Operator s Check ...
Страница 31: ...5 1 5 Performance Tests ...
Страница 40: ...5 10 Chapter5 Performance Tests Average Noise Level ...
Страница 41: ...6 1 6 Replaceable Parts ...
Страница 46: ...6 6 Chapter6 Replaceable Parts Parts Lists ...
Страница 47: ...7 1 7 Service ...
Страница 60: ...7 14 Chapter7 Service Replacement Procedure Figure 7 11 Regulator Parts and Wiring ...
Страница 64: ...7 18 Chapter7 Service Connector Inspection and Cleaning ...