4-4
Specifications and Regulatory Information
Specifications
20 GHz
≤
1 mV (0.5 mV typical)
Scale Factor (full scale is eight divisions)
Minimum
1 mV/div
Maximum
100 mV/div
dc Offset Range
±
500 mV
Nominal Impedance
50
Ω
Connector
3.5 mm (m)
Reflections
≤
5% for 30 ps rise time
Dynamic Range
±
400 mV relative to channel offset
Maximum Safe Input Voltage
16 dBm peak ac
±
2V dc
a. It is recommended that a user vertical calibration be performed after every 10 hours of continuous use or if the temperature has changed
by greater than 2
°
C from the previous vertical calibration.
b. Where
∆
T
cal
represents the temperature change in Celsius from the last user vertical calibration. Note that the temperature term goes to
zero upon execution of a vertical calibration.
c. Where
∆
Time
cal
represents the time since the last user vertical calibration. The uncertainty due to time typically stabilizes after 24 hours.
This term goes to zero upon execution of a vertical calibration.
Table 4-1. Agilent 83487A Electrical Channel Vertical Specifications (Continued)
Содержание 83487A
Страница 1: ...Agilent 83487A Optical Electrical Plug In Module User s Guide ...
Страница 8: ...viii General Safety Considerations ...
Страница 10: ......
Страница 25: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Страница 27: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Страница 28: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Страница 38: ......
Страница 67: ...4 9 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 68: ......
Страница 96: ......