3-5
Calibration Overview
Factory Calibrations
optional
Agilent 83480A, 54750A Service Guide
for more details about the
mainframe calibration, and the position of the rear-panel memory protect
switches.
C A U T I O N
To prevent access to the mainframe calibration switch, place a sticker over the
access hole to this switch.
C A U T I O N
Do not attempt a Mainframe calibration without consulting the
Agilent 83480A, 54750A Service Guide
.
A mainframe calibration should be performed on a periodic basis, annually, or
when the ambient operating temperature has changed by and remains 5
°
C dif-
ferent than the operating temperature at which the last mainframe calibration
was performed. To see how much the operating temperature has changed
since the last mainframe calibration and the date of the last mainframe calibra-
tion, check the Calibration status by pressing the following key sequence:
Utility
,
Calibrate
, and then
Cal status
on.
The temperature change is displayed at the top of the display as shown in the
following figure.
Figure 3-1. Current Frame
∆
Temp condition
Содержание 83487A
Страница 1: ...Agilent 83487A Optical Electrical Plug In Module User s Guide ...
Страница 8: ...viii General Safety Considerations ...
Страница 10: ......
Страница 25: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Страница 27: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Страница 28: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Страница 38: ......
Страница 67: ...4 9 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 68: ......
Страница 96: ......