Operation
36
Agilent 700 Series ICP Optical Emission Spectrometers User’s Guide
5
The Peltier will switch on automatically. If for some reason it has
been turned off, switch on the Peltier by choosing
Peltier on
from the
Analyze
menu.
NOTE
You can check if the Peltier is on by looking at the Instrument Setup window
Status page. It may take some time to reach operating temperature. It is
recommended that you wait three to five minutes before starting an analysis.
6
If you have accessories fitted, switch them on.
7
Switch on the laboratory exhaust system.
WARNING
Hot Surfaces and Noxious Fumes
Burn danger, inhalation hazard. The plasma emits heat, ozone and fumes
which can be hazardous. Always switch on the exhaust system before
lighting the plasma.
Preparing for Analysis
For detailed information on how to prepare your instrument for
analysis which includes setting up the peristaltic pump tubing and
turning on the plasma, see the ICP Expert II Help:
1
Click
Start > All Programs > ICP Expert II > Help
.
2
Expand
How to
and then click
Prepare the instrument for use
.