Agilent 700 Series ICP Optical Emission Spectrometers User’s Guide
Analysis Checklist
You need to complete the following steps in turn to measure a
sample/s. You will find information on each step in this chapter.
Turn on the instrument and software
Prepare for analysis
Calibrate the detector
Align the torch
Set operating conditions
Create/open a worksheet
Develop a method
Run samples
Print a report
Turning On the Instrument and Software
Before starting the system, carefully read the Safety practices and
hazards section at the front of this manual and ensure that the
laboratory is set up according to the details specified in the Site
Preparation Guide.
Turning On the Instrument for the First Time (or From Shutdown
Mode)
To turn on the instrument for the first time or from Shutdown
mode:
1
Check that all tubing on the torch, spraychamber, nebulizer and
peristaltic pump is correctly connected.
2
Close the torch compartment door, ensuring that the locking
lever is fully latched.
3
Switch on the computer, monitor and printer.
4
Switch on the water cooler.
5
Turn on the gas supply at the cylinder.