Operation
Agilent 700 Series ICP Optical Emission Spectrometers User’s Guide
35
6
Plug the cable into the wall socket and set the switch to ‘On’.
7
To turn on the spectrometer, set the RF supply circuit breaker
located on the right rear side of the instrument to the ‘On’ (up)
position.
8
Set the instrument power switch, located at the front left of the
instrument, to ‘
|
’ (the green power indicator should light).
NOTE
Both the RF supply circuit breaker and instrument power switch must be
switched on for the instrument to work.
9
Switch on the laboratory exhaust system.
WARNING
Hot Surfaces and Noxious Fumes
Burn danger, inhalation hazard. The plasma emits heat, ozone and fumes,
which can be hazardous. Always switch on the exhaust system before
lighting the plasma.
10
To start the ICP Expert II software, click
Start > All
Programs >
ICP Expert II > ICP Expert II
. The Main Index window will
appear.
Turning On the Instrument from Standby Mode
To turn on the instrument from Standby mode:
1
Check that all tubing on the torch, spraychamber, nebulizer and
peristaltic pump is correctly connected.
2
Close the torch compartment door, ensuring that the locking
lever is fully latched.
3
Switch on the monitor and printer (if they are off).
4
Switch on the water cooler (if it is off).