5 - Front Panel Operation
44
V or A HIGH
V or A MAX
V or A MIN
100 millisecond
V or A LOW
acquisition time
46.8 microsecond
sampling rate
NOTE:
coincide with the actual maximum
or minimum point of the waveform.
Measurement samples may not
Figure 5-5. Front Panel Pulse Measurement Parameters (Agilent 66332A only)
Use the Meter menu for making front panel measurements:
Action
Display
1.
For current measurements, press
Shift, Input
. Then press
Ì
until you
obtain the CURR:RANG AUTO command. Press
Enter
to activate
autoranging. Two other selections are also available. Select the High
range when measuring currents above 20 mA. Select the Low range for
improved resolution when measuring currents below 20 mA.
CURR:RANG AUTO
2.
For output waveform measurements, press
Shift, Input
. Then press
q
until you obtain the CURR:DET command. Check to make sure that the
ACDC current detector is selected. This provides the best accuracy for
waveform measurements. Only select the DC current detector if you are
making dc current measurements and you require a dc measurement
offset better than 1mA on the High current measurement range.
CURR:DET ACDC
Note:
In the Low current measurement range, the current detector is fixed at DC. With the
current detector in dc, accurate current measurements cannot be made on waveforms
with frequency contents over a few kHz.
3.
On the Function keypad press
Meter
and press
q
repeatedly to
access the following measurement parameters:
♦
dc voltage and current
♦
peak voltage
1
♦
minimum voltage
1
♦
high level of a voltage pulse waveform
1
♦
low level of a voltage pulse waveform
1
♦
rms voltage
1
♦
peak current
1
♦
minimum current
1
♦
high level of a current pulse waveform
1
♦
low level of a current pulse waveform
1
♦
rms current
1
1
Agilent 66332A only
<reading>V <reading>A
<reading>V MAX
<reading>V MIN
<reading>V HIGH
<reading>V LOW
<reading>V RMS
<reading>A MAX
<reading>A MIN
<reading>A HIGH
<reading>A LOW
<reading>A RMS