3 - Troubleshooting
32
"F" From Sheet 9
Disable the OV
protection by sending
the command
"DIAG:SCR 0"
CV annunciator
on?
Voltmeter reads
+OL?
No
Disabled
annunciator on?
Gated +/-15 Biases
at zero?
Yes
Check U311/U312
Gated Bias circuits
Check Series
Regulator circuits,
U309, Q303, Q308 etc.
No
Check Vmon amplifier,
U322B (Vmon = 0 to
4.7V for 0 to full scale
output voltage)
No
Check U316A (gain of
-1) and U322A circuits
Yes
"CONTROL"
U321B-7
< 0V ?
Yes
Check U321B circuit
No
Check Series
Regulator circuits,
U309, Q303, Q308 etc.
Yes
PM_INHIBIT @
R380 15V
?
No
Yes
Yes
Replace A2 Interface
board and go back to
beginning. Probable
2nd problem.
No
Figure 3-1 Sheet 10. Output Held High