Verification and Performance Tests - 2
19
CC Noise (PARD)
Periodic and random deviations (PARD) in the output combine to produce a residual ac current, as well, as an ac
voltage superimposed on the dc output. Constant current (CC) PARD is specified as the rms output current in a
frequency range 20 Hz to 20 Mhz with the supply in CC operation.
a.
Turn off the supply and connect the load, monitoring resistor, and rms voltmeter as shown in Figure 2-1a. The
Current Monitoring resistor may have to be substituted by one with a higher resistance and power rating, such
as a 1 ohm 5 W current shunt in series with a 9 ohm resistor, to get the RMS voltage drop high enough to
measure with the RMS voltmeter. Leads should be as short as possible to reduce noise pick-up. An electronic
load may contribute ripple to the measurement so if the RMS noise is above the specification a resistive load
may have to be substituted for this test.
b. Check the test setup for noise with the supply turned off. Other equipment (e.g. computers, DVMs, etc.) may
affect the reading.
c.
Turn on the supply and program the current to full scale and the output voltage to the maximum programmable
value.
d. The output current should be at the full scale rating with the CC annunciator on.
e.
Divide the reading on the rms voltmeter by the monitor resistance to obtain rms current. It should not exceed the
values listed in the performance test record card under CC NOISE (RMS).
Performance Test Equipment Form
Test Facility:
_________________________
Report Number ________________________
____________________________________
Date _________________________________
____________________________________
Customer _____________________________
____________________________________
Tested By ____________________________
Model ______________________________
Ambient Temperature (C) ________________
Serial No. ____________________________
Relative Humidity (%) ___________________
Options _____________________________
Nominal Line Frequency __________________
Firmware Revision ____________________
Special Notes:
Test Equipment Used:
Description
Model No.
Trace No.
Cal. Due Date
AC Source
_________________
_________________
_________________
DC Voltmeter
_________________
_________________
_________________
RMS Voltmeter
_________________
_________________
_________________
Oscilloscope
_________________
_________________
_________________
Electronic Load
_________________
_________________
_________________
Current Shunt
_________________
_________________
_________________
_________________
_________________
_________________
Содержание 6612B
Страница 8: ......
Страница 25: ...Troubleshooting 3 25 Figure 3 1 Sheet 3 Main Flowchart...
Страница 46: ......
Страница 49: ...Principles of Operation 4 49 Figure 4 1 A2 A3 Block Diagram...
Страница 51: ...Principles of Operation 4 51 Figure 4 2 A1 Block Diagram...
Страница 54: ......
Страница 57: ...Replaceable Parts 5 57 Figure 5 1 Mechanical Parts ldentification...
Страница 67: ...Figure 6 1 A1 Board Component Locations...
Страница 68: ...Figure 6 2 A4 Board Component Locations...
Страница 69: ...Figure 6 3 A1 Board schematic sheet 1...
Страница 70: ...Figure 6 3 A1 Board schematic sheet 2...
Страница 71: ...Figure 6 3 A1 Board schematic sheet 3...
Страница 72: ......