8 – Language Dictionary
138
*RST
This command resets the dc source to a factory-defined state as defined in the following table. *RST also
forces an ABORt command.
Table 8-7. *RST Settings
CAL:STAT
OFF
[SOUR:]CURR2 10% of MAX*
DIG:DATA 0
[SOUR:]CURR2:TRIG 10% of MAX*
DISP:STAT
ON
[SOUR:]CURR:PROT:STAT OFF
DISP:MODE
NORM
[SOUR:]LIST:COUN
0
DISP:TEXT “ “
[SOUR:]VOLT 0
INIT:CONT
OFF
[SOUR:]VOLT:TRIG 0
INST:COUP:OUTP:STAT
ALL
[SOUR:]VOLT2 0
NONE (option 521 units)
[SOUR:]VOLT2:TRIG 0
OUTP
OFF
[SOUR:]VOLT:PROT MAX*
OUTP:DFI
OFF
[SOUR:]VOLT:PROT:STAT ON
OUTP:DFI:SOUR
OFF
TRIG:ACQ:COUN:CURR 1
OUTP:PROT:DEL
.08
TRIG:ACQ:COUN:VOLT 1
OUTP:PROT:STAT
OFF
TRIG:ACQ:HYST:CURR 0
OUTP:TYPE
LOW
TRIG:ACQ:HYST:DVM 0
SENS:CURR:RANG MAX
TRIG:ACQ:HYST:VOLT 0
SENS:CURR:DET ACDC (all but 66311A)
TRIG:ACQ:LEV:CURR 0
DC (66311A only)
TRIG:ACQ:LEV:DVM 0
SENS:FUNC VOLT
TRIG:ACQ:LEV:VOLT 0
SENS:SWE:OFFS:POIN
0
TRIG:ACQ:SLOP:CURR POS
SENS:SWE:POIN
2048
TRIG:ACQ:SLOP:DVM POS
SENS:SWE:TINT
15.6
µ
s
TRIG:ACQ:SLOP:VOLT POS
[SOUR:]CURR 10% of MAX*
TRIG:ACQ:SOUR INTERNAL
[SOUR:]CURR:TRIG 10% of MAX*
TRIG:TRAN:SOUR BUS
Command Syntax
*RST
Parameters
None
Related Commands
*PSC *SAV
*SAV
This command stores the present state of the dc source to the specified location in non-volatile memory.
Up to 4 states can be stored. If a particular state is desired at power-on, it should be stored in location 0.
It will then be automatically recalled at power turn-on if OUTPut:PON:STATe is set to RCL0. Use
*RCL to retrieve instrument states.
Command Syntax
*SAV <NRf>
Parameters
0 | 1 | 2 | 3
Example
*SAV 3
Related Commands
*RCL *RST
CAUTION:
*SAV causes a write cycle to nonvolatile memory. Nonvolatile memory has a finite
maximum number of write cycles. Programs that repeatedly cause write cycles to
nonvolatile memory can eventually exceed the maximum number of write cycles and
cause the memory to fail.