3 - Installation
44
Dc sources may be connected to the GPIB interface in series configuration, star configuration, or a
combination of the two, provided the following rules are observed:
♦
The total number of devices including the GPIB interface card is no more than 15.
♦
The total length of all cables used is no more than 2 meters times the number of devices connected
together, up to a maximum of 20 meters. (Refer to table 2-2 for a list of GPIB cables available from
Agilent Technologies.)
♦
Do not stack more than three connector blocks together on any GPIB connector.
♦
Make sure all connectors are fully seated and the lock screws are firmly finger-tightened.
RS-232 Interface
Agilent 66111A and 66311B/D dc sources have an RS-232 programming interface, which is activated by
commands located in the front panel
Address
menu. All applicable SCPI and COMPatibility commands
are available through RS-232 programming. When the RS-232 interface is selected, the GPIB interface is
disabled.
The RS-232 connector is a DB-9, male connector. Adapters are available to connect the dc source to any
computer or terminal with a properly configured DB-25 connector (see Table 2-2).
1 2 3 4 5
6 7 8 9
Figure 3-12. RS-232 Connector
Pin
Input/Output
Description
1
-
no connection
2
Input
Receive Data (RxD)
3
Output
Transmit Data (TxD)
4
Output
Data Terminal Ready (DTR)
5
Common
Signal ground
6
Input
Data Set Ready (DSR)
7
Output
Request to Send (RQS)
8
Input
Clear to Send (CTS)
9
-
no connection