4

Fan Bracket Assistance Tool

3

While prying down, use thumb to push bracket into position.

Содержание 54845-68803

Страница 1: ...s Fan Safety Shield Instructions Follow these instructions to install and use the Fan Safety Shield W A R N I N G Risk of Injury Use caution when working around the Infiniium cooling fan s with the instrumentsleeves removed The cooling fan s exposed on one side can be hazardous When the cabinet sleeves are removed install the Fan Safety Shield supplied with this kit to protect your fingers from th...

Страница 2: ...om cover If necessary refer to the Replacing Assemblies procedures in the Infiniium 54845B 46B Oscilloscope Service Manual 4 Clipthefanshield over the outsideofthechassisasshowninthepicture below 5 You may now service the ATX Oscilloscope with power applied 6 You must remove safety shield before re installing sleeve and bottom cover The safety shield contains a raised ridge to prevent against this...

Страница 3: ... removal and installation of the ATX frame fan brackets Removing Fan Bracket 1 Insert the assistance tool as shown in the picture below 2 With tool in place pry upwards to remove the bracket Installing Fan Bracket 1 Using thumb push bracket up against chassis 2 Insert assistance tool as shown below ...

Страница 4: ...4 Fan Bracket Assistance Tool 3 While prying down use thumb to push bracket into position ...

Страница 5: ...niium oscilloscope 2 Connect a mouse and a keyboard to the oscilloscope 3 Insert the service disk in the floppy drive and turn on the oscilloscope 4 The oscilloscope will boot into the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Highlight Programs select Accessories and then select System Tools Click the Scan Disk icon 7 In the ScanDisk applicatio...

Страница 6: ...nnect a mouse and a keyboard to the oscilloscope 3 Insert the service disk in the floppy drive and turn on the oscilloscope 4 The oscilloscope will boot into the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Highlight Programs and click the Infiniium Scope icon ...

Страница 7: ...the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Open the Programs group and click on the Infiniium Scope icon 7 When the oscilloscope application comes up click on the mouse icon in the upper right corner of the screen 8 Click on the Utility pull down menu click on Selftest In the Selftest screen click on Select Service Extensions 9 Click on the S...

Страница 8: ...nel you are changing 3 Under Set Value enter the corrected attenuator value using the keyboard or the drop down keypad then press the button labeled Apply to Set Value to Actuation Count 4 Repeat for each channel you need to change Agilent Technologies Printed in Malaysia Agilent Part Number 54845 92001 Printed January 2002 ...