Fan Safety Shield Instructions
Installing The Fan Safety Shield
1
Turn off the power.
2
Disconnect the instrument power cord and any network or printer
cables.
3
Remove the instrument sleeve and bottom cover. If necessary, refer to
the “Replacing Assemblies” procedures in the Infiniium
54845B/46B Oscilloscope Service Manual.
4
Clip the fan shield over the outside of the chassis as shown in the picture
below.
5
You may now service the ATX Oscilloscope with power applied.
6
You must remove safety shield before re-installing sleeve and bottom
cover. The safety shield contains a raised ridge to prevent against this.