Setting Up the Oscilloscope
To connect oscilloscope probes
1-21
3
To disconnect the probe, push the small latch on top of the probe
connector to the left, then pull the connector body away from the front
panel of the oscilloscope without twisting it.
Figure 1-14
Disconnecting the Oscilloscope Probe
C A U T I O N
Do not attempt to twist the snap-on probes on or off the oscilloscope’s BNC
connector. Twisting the probe connector body will damage it.
C A U T I O N
Do not exceed the maximum input voltage rating! The maximum input voltage
for 50
Ω
inputs is 5 Vrms, CAT I. Maximum voltage for the 54810A/15A/20A/
25A at 1 M
Ω
input impedance is ±250V (dc + ac) [ac < 10 kHz], CAT I; for the
54835A and 54845A it is ±100V (dc + ac) [ac < 10 kHz], CAT I.
!