3- 20
Triggering the Oscilloscope
To use CAN triggering
2 Press the
Baud
softkey to set the CAN signal baud rate to match your
CAN bus signal.
The CAN baud rate can be set to:
The default baud rate is 1 Mb/s
3 Press the
Trigger:
softkey to select the trigger condition.
The
SOF
(Start of Frame) is the only trigger condition available. The CAN trigger
will trigger on the Start of Frame bit of a Data frame, Remote Transfer Request
(RTR) frame, or an Overload frame.
4 Press the
Signal
softkey to set the type and polarity of the CAN signal.
This also automatically sets the channel label for the source channel. The
channel shown in the
Source
softkey can be connected as follows:
Dominant high signal:
•
CAN_H
- the actual CAN_H differential bus signal.
Dominant low signals:
•
CAN_L
- the actual CAN_L differential bus signal.
•
Rx
- the Receive signal from the CAN bus transceiver.
•
Tx
- the Transmit signal to the CAN bus transceiver.
•
Differential
- the CAN differential bus signals connected to an analog
source channel using a differential probe.
5 Press the
Source
softkey to select the channel connected to the CAN
signal line.
If the Signal softkey is set to Differential, connect an analog source channel to
the CAN differential bus signal using a differential probe.
As you press the
Source
softkey (or rotate the Entry knob on mixed-signal
oscilloscopes), the
CAN
label for the source channel is automatically set and the
channel you select is shown in the upper-right corner of the display next to
"
CAN
".
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob. Select
Threshold
in the
D7 Thru D0
or
D15 Thru D8
menu to set the
level for digital channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
10 kb/s
50 kb/s
100 kb/s
500 kb/s
20 kb/s
62.5 kb/s
125 kb/s
800 kb/s
33.3 kb/s
83.3 kb/s
250 kb/s
1 Mb/s
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...