MegaZoom Concepts and Oscilloscope Operation
To assign a predefined label to a channel
To assign a predefined label to a channel
1 Press the
Label
key.
2 Press the
Channel
softkey, then turn the Entry knob or successively press
the softkey to select a channel for label assignment.
The figure above shows the list of channels and their default labels. The channel
does not have to be turned on to have a label assigned to it.
3 Press the
Library
softkey, then turn the Entry knob or successively press
the
Library
softkey to select a predefined label from the library.
4 Press the
Save New Label
softkey to assign the label to your selected
channel.
5 Repeat the above procedure for each predefined label you want to
assign to a channel.