6-15
Utilities
To perform service functions
User Cal Status
Pressing the
User Cal Status
softkey displays the following summary results of
the previous User Cal.
Results:
User Cal date:
DT since last User Cal:
Failure:
Comments:
Self test
Pressing the
Start Self Test
softkey performs a series of internal procedures to
verify that the oscilloscope and any attached module are operating properly.
It is recommended you run Self Test:
• after experiencing abnormal operation.
• for additional information to better describe an oscilloscope failure.
• to verify proper operation after the oscilloscope has been repaired.
Successfully passing Self Test does not guarantee 100% of the oscilloscope's
functionality. Self Test is designed to provide an 80% confidence level that the
oscilloscope is operating properly.
About Oscilloscope
Pressing the
About Oscilloscope
softkey displays information about your
oscilloscope model number, serial number, software revision, ROM revision, and
option module attached.
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...