3- 10
Triggering the Oscilloscope
External Trigger Input
2 Press the
External
softkey to display the external trigger probe menu.
Probe Attenuation
Turn the Entry knob to set the attenuation factor
displayed in the
Probe
softkey for the connected probe. The attenuation factor
can be set from 0.1:1 to 1000:1 in a 1-2-5 sequence.
For 2-channel oscilloscopes that have the Ext Trigger connector on the front
panel, when you connect an AutoProbe self-sensing probe, the oscilloscope will
automatically configure your probe to the correct attenuation factor.
The probe correction factor must be set properly for measurements to be made
correctly.
Probe Units
Press the
Units
softkey to select the proper measurement unit
for the connected probe. Select
Volts
for a voltage probe and select
Amps
for a
current probe. Measurement results, channel sensitivity, and trigger level will
reflect the measurement units you have selected.
Input Impedance
The 54620-series oscilloscope external trigger input
impedance is always 1 M
Ω
. The 54640-series oscilloscope external trigger input
impedance can be set to either
1M Ohm
or
50 Ohm
by pressing the
Imped
softkey.
•
50 Ohm
mode matches 50-ohm cables commonly used in making high
frequency measurements. This impedance matching gives you the most
accurate measurements since reflections are minimized along the signal
path.
•
1M Ohm
mode is for use with probes and for general-purpose measurements.
The higher impedance minimizes the loading effect of the oscilloscope on
the circuit under test.
For 2-channel oscilloscopes that have the Ext Trigger connector on the front
panel, when you connect an AutoProbe self-sensing probe, the oscilloscope will
automatically configure your probe to the correct impedance.
Attenuation
factor
Measurement
units
Input impedance
(54640-series)
Return to
previous menu
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...