Remote Operation
66
Table 5-1. Power Supply Commands (continued)
Command
Header
Output
Channel
Data Range
(Fig. 5-2)
Syntax
Query present hardware error
ERR?
---
See Table 5-8
Q1
Query fault register
FAULT?
1, 2, 3, 4
---
Q2
Query the model number of supply
ID?
---
---
Q1
Program the I DAC in counts
IDAC
1, 2, 3, 4
See Service Manual
C4
Query setting of I DAC in counts
IDAC?
1, 2, 3, 4
---
Q2
Send data to calibrate I circuits
IDATA
1, 2, 3, 4
See Table 5-4
C5
Sets output to high I cal. value
IHI
1, 2, 3, 4
See Table 5-4
C3
Sets output to low I cal. value
ILO
1, 2, 3, 4
See Table 5-4
C3
Query measured I output
IOUT?
1, 2, 3, 4
See Table 5-4
Q2
Used to Calibrate I readback circuits
IRDAT
1, 2, 3, 4
See Table 5-4
C5
Set output to + I readback high cal value
IRHI
1, 2, 3, 4
See Table 5-4
C3
Set output to - I readback high cal value
IRHN
1, 2, 3, 4
See Table 5-4
C3
Set output to - I readback low cal value
IRLN
1, 2, 3, 4
See Table 5-4
C3
Set output to + I readback low cal value
IRLO
1, 2, 3, 4
See Table 5-4
C3
Set full scale current range of output
IRSET
1, 2, 3, 4
See Table 5-4
C4
Query full scale current range
ISET?
1, 2, 3, 4
See Table 5-4
Q2
Set current of an output
ISET
1, 2, 3, 4
See Table 5-4
C4
Query current of an output
ISET?
1, 2, 3, 4
See Table 5-4
Q2
Increase or decrease output current by value
ISTEP
1, 2, 3, 4
See Table 5-4
C4
Select which output will be metered
METER
---
1-4
C2
Query which output is being metered
METER?
---
---
Q1
Send data to calibrate -I readback
NIDAT
1, 2, 3, 4
See Table 5-4
C5
Enable overcurrent protection
OCP
1, 2, 3, 4
0,1 (OFF,ON)
C4
Query if OCP is enabled
OCP?
1, 2, 3, 4
---
Q2
Reset overcurrent protection
OCRST
1, 2, 3, 4
---
C3
Turn output on or off
OUT
1, 2, 3, 4
0,1 (OFF,ON)
C4
Query if output is on or off
OUT?
1, 2, 3, 4
---
Q2
Perform overvoltage calibration
OVCAL
1, 2, 3, 4
See Table 5-4
C3
Reset overvoltage circuit
OVRST
1, 2, 3, 4
---
C3
Set overvoltage trip value
OVSET
1, 2, 3, 4
See Table 5-4
C4
Query overvoltage trip value
OVSET?
1, 2, 3, 4
See Table 5-4
Q2
Enable power on service request
PON
---
0,1 (OFF,ON)
C2
Query if PON is enabled
PON?
1, 2, 3, 4
---
Q1
Recall voltage and current settings
RCL
---
0-10
C2
Set readback DAC to a number of counts
RDAC
1, 2, 3, 4
See Service Manual
C4
Query readback DAC count setting
RDAC?
1, 2, 3, 4
---
Q2
Query revision date of ROM
ROM?
---
Q1
Query revision date of secondary ROM
SROM?
1, 2, 3, 4
---
Q2
Set causes for generating a service request
SRQ
---
01,2,3
C2
Query causes which will generate an SRQ
SRQ?
---
---
Q1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 47513-7
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 25: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 31: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 45: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 63: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 117: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 121: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...