Check
13,
Transformer
Secondary
V
oltage
1.
Turn
o
the
4339B .
2.
Disconnect
the
transformer
secondary
connector
from
connector
\J13"
on
the
A1
Main
Board
Assembly
.
Refer
to
Figure
4-14
for
the
connector
location.
3.
Turn
the
4339B
on.
4.
Using
a
multimeter
,
check
if
the
transformer
secondary
voltage
is
between
13
V
and
17
V
ac
.
5.
If
passed,
replace
the
A1
Main
Board
Assembly
,
otherwise
replace
the
transformer
assembly
.
Figure
4-14.
Transformer
Secondary
V
oltage
Check
4-18
T
roubleshooting