A1
Main
Board
Removal
Removal
Procedure
1.
Remove
the
A2
CPU
Board
Assembly
while
referring
to
\A2
CPU
Board
Assembly
Removal".
2.
Disconnect
the
following
cable
assemblies
from
the
A1
assembly
.
a.
F
our
cable
assemblies
from
the
front
panel
b.
The
cable
assembly
from
the
transformer
c.
Two
cable
assemblies
from
the
DC-DC
Converter
3.
Disconnect
the
four
optical
ber
cables
which
are
connected
to
the
A3
Ammeter
Board
Assembly
4.
Disconnect
the
four
optical
ber
cables
from
the
A3
Ammeter
Board
Assembly
Caution
T
o
avoid
damaging
on
the
optical
ber
cable
caused
by
bending
or
other
mechanical
stress
,
remove
all
the
ber
cables
from
the
4339B
when
replacing
either
the
A1
board
or
the
A3
board.
5.
Remove
the
cable
clamp
on
the
shield
case
which
surrounds
the
A3
board.
6.
Remove
the
nut
which
fastens
the
Ext
Trigger
connector
on
the
rear
panel.
7.
Remove
the
screw
which
secures
the
A1
board
to
the
chassis
side
.
8.
Remove
the
four
studs
,
that
secure
the
A1
assembly
to
the
chassis
,
by
rotating
the
studs
with
a
at
bladed
screwdriver
.
9.
Remove
the
three
screws
that
secure
the
A1
assembly
to
the
chassis
.
10.
Remove
the
A1
assembly
from
the
chassis
.
A2
CPU
Board
Assembly
Removal
Removal
Procedure
1.
Remove
the
cover
while
referring
to
\Cover
Removal".
2.
Disconnect
the
four
atcable
assemblies
and
the
wire
assembly
from
the
A2
board.
3.
Remove
the
four
screws
that
secure
the
A2
board
to
the
stud.
4.
Remove
the
A2
board
from
the
chassis
.
Note
The
ROM
and
EEPROM
on
the
A2
assembly
may
be
transferred
to
the
new
assembly
to
minimize
the
repair
time
and
parts
cost.
Refer
to
\Replacing
the
A2
CPU
Board
Assembly"
for
more
information.
3-12
Assembly
Replacement