
Figure
3-2.
Cable
Assembly
Connection
for
Interlock
Connector
Triaxial
Connector
Removal
Caution
When
accessing
the
traiaxial
connector
,
use
lint
free
gloves
to
avoid
contaminating
the
connector
.
Agilent
T
echnologies
supplies
lint
free
gloves
(Agilent
PN
9300-0163).
Removal
Procedure
1.
Remove
the
cover
while
referring
to
\Cover
Removal".
2.
Disconnect
the
two
atcable
assemblies
and
the
wire
assembly
from
the
front
side
of
the
A2
board.
3.
Remove
the
screw
which
fastens
the
cable
assembly
on
the
front
side
of
the
A3
Ammeter
Board
Assembly
.
4.
Remove
the
screws
which
fastens
the
cable
assembly
on
the
left
front
side
of
the
chassis
.
5.
Remove
the
side
trim
label
from
both
sides
of
the
front
bezel.
6.
Remove
the
four
screws
that
secure
the
front
panel
bezel
to
the
chassis
from
both
sides
.
7.
Place
the
front
panel
assembly
,
so
that
the
front
panel
faces
down.
8.
Disconnect
the
cable
assembly
which
includes
a
core
from
the
front
side
of
the
A3
Ammeter
board
assembly
.
9.
Remove
the
core
holder
which
holds
the
cable
assembly
disconnected
in
the
last
procedure
,
by
unscrewing
the
screw
which
secures
the
holder
from
the
front
side
.
Unsolder
the
diodes
from
the
sleeve
of
the
triaxial
connector
.
10.
Remove
the
heat
shrink
tube
from
the
triaxial
connector
.
11.
Unsolder
the
sleeve
from
the
triaxial
connector
.
12.
Unsolder
the
coaxial
cable
from
the
triaxial
connector
.
13.
Unscrew
the
nut
which
fastens
the
triaxial
connector
to
the
sub
panel.
Assembly
Replacement
3-15