14.
Remove
the
triaxial
connector
.
Note
When
soldering
the
cable
assembly
to
the
triaxial
connector
,
refer
to
Figure
3-3
and
carefully
solder
the
cable
to
the
connector
so
that
the
center
conductor
and
the
outer
shield
are
not
shorted
to
each
other
.
Figure
3-3.
Cable
Assembly
Connection
for
Triaxial
Connector
BNC
Connector
Removal
Removal
Procedure
Caution
When
accessing
the
BNC
connector
,
use
lint
free
gloves
to
avoid
contaminating
the
connector
.
Agilent
T
echnologies
supplies
lint
free
gloves
(Agilent
PN
9300-0163) .
1.
Remove
the
cover
while
referring
to
\Cover
Removal".
2.
Disconnect
the
two
atcable
assemblies
and
the
wire
assembly
from
the
front
side
of
the
A2
board.
3.
Remove
the
screw
which
fastens
the
cable
assembly
on
the
front
side
of
the
A3
Ammeter
Board
Assembly
.
4.
Remove
the
screws
which
fastens
the
cable
assembly
on
the
left
front
side
of
the
chassis
.
5.
Remove
the
side
trim
label
from
both
sides
of
the
front
bezel.
6.
Remove
the
four
screws
that
secure
the
front
panel
bezel
to
the
chassis
from
both
sides
.
7.
Place
the
front
panel
assembly
,
so
that
the
front
panel
faces
down.
8.
Remove
the
heat
shrink
tube
from
the
BNC
connector
being
replaged.
9.
Unsolder
the
cable
from
the
BNC
connector
being
replaced.
10.
Unscrew
the
nut
which
connects
the
BNC
connector
to
the
sub
panel.
11.
Remove
the
BNC
connector
.
3-16
Assembly
Replacement