Agilent E5025 Cryogen Monitor
5
400/54 Premium Shielded NMR Magnet System Product Overview
71
•
100% corresponds to full
•
0% means that the level has dropped to the refill level
for the magnet
The accuracy of helium level readings is approximately 2mm
(better than 1%). Below the refill level, the display outputs a
negative reading which moves to –100% as the level drops to
the bottom of the helium level probe. If the level falls below
the refill level (0%) then the helium display flashes to
indicate a low level and that a fill is required.
The unit is preprogrammed to read the helium level probes
every eight hours at the internal clock preset times of 01:00,
09:00 and 17:00. As a stand- alone unit, the internal clock is
factory set to GMT timezone. If connected to a console (with
the appropriate software), then the internal clock is changed
to the time registering on the console.
Users of the E5025 have access to a number of set- up items
via the front panel buttons. Two menu levels provide a top
level “user” menu and a secondary level “set- up” menu. Each
menu is traversable in the down direction only.
Note when pressing the buttons on the display:
•
A finger on or off button delay of about 1/3 second is
implemented, therefore do not press buttons too quickly.
•
A flashing display indicates an error or alarm situation.
There is a regular period, every five or so seconds, when
the flashing is held off for a time of 3 seconds. If the
display is in flash mode, use the three- second nonflashing
period to press buttons.
•
Buttons are scanned several times per second. However it
is possible the press can be missed when the instrument
is carrying out internal operations. In this situation,
release the buttons and try again.
•
All commands have an automatic time- out of
approximately 30 seconds, after which the unit returns to
the cryogen display mode. Any value chosen for
adjustment but not “entered” is ignored and the original
value remains. Likewise any action is abandoned.
Содержание 400/54 Premium Shielded NMR Magnet System
Страница 1: ...Agilent Technologies Agilent 400 54 Premium Shielded NMR Magnet System System Overview ...
Страница 6: ...4 400 54 Premium Shielded NMR Magnet System Product Overview ...
Страница 32: ...30 400 54 Premium Shielded NMR Magnet System Product Overview 2 Installation and Operation Safety ...
Страница 36: ...34 400 54 Premium Shielded NMR Magnet System Product Overview 3 Maintenance Figure 2 Magnet orientation ...
Страница 56: ...54 400 54 Premium Shielded NMR Magnet System Product Overview 3 Maintenance ...
Страница 66: ...64 400 54 Premium Shielded NMR Magnet System Product Overview 4 Reference ...
Страница 99: ......
Страница 100: ...Agilent Technologies Agilent Technologies Inc Printed in USA April 2012 9100380100 9100380100 ...